Model 100
- Superior HPLC Detection to Replace or Complement your Current Detection
- Sensitive Detection for a Wide Variety of Chemical Classes and Structures
- Patented Thermo-Split™ Technologyfor Precise Vapor Phase Control (Patent No. US 7,290,723 B1
- Easy to Use, fast start up
- Low Cost of Operation with no consumable parts
Specifications
Dimensions | 9.8” W x 18” D x 11.5” H |
Weight | 25 lbs |
Display | 2 Line x 20 Character per line VFD |
User Interface | Two multi-function buttons |
Evaporative Zone Temperature | 30°C |
Thermo-Split™ Chamber Temperature | 60°C |
Liquid Flow Rate | 0.2mL/min to 3mL/min |
Gas Requirements | 60 – 80 psi Nitrogen or other inert gas |
Gas Consumption | ~ 2.5 SLPM |
Operating Conditions | Intended for indoor use only, 60°F to 85°F and < 90% R.H. non condensing |
Electrical Requirements | Nominal 120 VAC, 50/60 Hz or Nominal 240 VAC, 50/60 Hz; 600 watts |
Wetted Materials | Stainless steel, glass, anodized aluminum, Teflon™ |
Light Source | 670 nm Laser Diode, <5mW |
Detector | Hermetically sealed photo-diode/operational amplifier |
Output Signal | 0 – 5 VDC |
Interface | RS232, Contact Closure |