Principle | FT-IR microscope with cassegrain optical system |
Measurement Method | Transmittance / Reflectance measurement |
Standard Detector | IRT-5100: DLATGS detector (7800-400 cm -1 ) IRT-5200 & IRT-7100: Single mid-band MCT (7800-600 cm -1 ) IRT-7200: Linear array MCT (7000-650 cm -1 x 16 element) Single mid-band MCT (7800-600 cm) -1 |
Detector exchange | Dual detector capability (software controlled), user exchangeable single element detectors are available as an option. |
Optional Detectors | Single element detector (IRT-5100): – Narrow-band: MCT (5000-750 cm -1 ) – Mid-band: MCT (7800-600 cm -1 ) – Wide-band: MCT (7800-450 cm -1 ) – InSb: (15000-1850 cm -1 ) – InGaAs detector: (12000-4000 cm -1 ) Single element detector (IRT-5200, IRT-7100, & IRT-7200): Linear array detector (IRT-5200, IRT-7100): Linear array detector (IRT-7200): |
S/N Ratio | Single element detector (IRT-5100): – 1000:1 (Aperture size 300 μm 2 , resolution 4 cm -1 , 1 min. acquisition, near 2200 cm -1 , p-p) Single element detector (IRT-5200, IRT-7100, & IRT-7200): Linear array detector (IRT-7200): |
Microscope objectives | IRT-5100, IRT-5200, & IRT-7100: – Cassegrain: 16 x 32 x or 10 x – Automatic objective recognition function (standard) – Up to four objectives can be selected by the software. IRT-7200: |
Condenser mirror | IRT-5100, IRT-5200, & IRT-7100: – Cassegrain: 16 x 32 or 10 x (manual exchange) – Automatic condenser mirror recognition function (standard) IRT-7200: |
Condenser mirror compensation | Standard auto-compensation function |
Aperture | PC-controlled vertical horizontal adjustment and angle of rotation |
Sample Storage | Standard (IRT-5100 & IRT-5200): – Manual stage with fine adjustment (Movable distance: X: 70, Y: 50, Z: 20 mm) Standard (IRT-7100 & IRT-7200): Option (IRT-5100 & IRT-5200): Option (IRT-7100 & IRT-7200): |
Auto focus | Option (IRT-5100 & IRT-5200) Standard (IRT-7100 & IRT-7200) |
IQ Mapping | Standard (IRT-7100 & IRT-7200) |
Sample Observation | High-resolution, 3 Megapixel CMOS camera with a 3X optional zoom function (standard) IQ Monitor (simultaneous sample measurement and observation) and auto illumination function (standard) 5.7 inch integrated color LCD display (option), Binocular (option) |
Observation Options | Visible polarization observation, Differential interference contrast observation, Fluorescence observation Refractive objectives: 10×, 20× |
Microscope Objectives | Cassegrain 16× – standard; 10×, 32× – optional Automatic objective recognition function is standard. Up to four objectives can be selected by the software. |
ATR Measurement (option) | “Clear-View” ATR objective (ATR-5000-SS/SD/SG) *1 , conventional ATR objective (ATR-5000-D/Z/G) *1 , Stage-mounted micro ATR using transmittance light path (ATR-5000-TPZ) |
Grazing Angle Measurement (option) | Cassegrain grazing angle objective (RAS-5000) *2 |
Purge | Sample area purge case is optional as an option. |
Integrated Control Panel | Transmittance/Reflection switching with indicator; detector indicator; objective selection/indicator; open/close and rotation of aperture; auto-compensation of condenser mirror; visible illumination adjustment |
Dimension | IRT-5100 & IRT-5200: 587 (H) × 302 (W) × 695 (D) mm, 54 kg IRT-7100 & IRT-7200: |
Power Consumption | IRT-5100 & IRT-5200: AC 100-240 V, 50/60 Hz, max. 60 VA IRT-7100 & IRT-7200: |
*1 Pressure sensor (PRS-M-5000, PRS-A-5000) is required.
*2 Infrared polarizer (PL-IR-5000, PL-IR-7000) is required.