Temperature-control-kinetics

STR-773 Water thermostatted cell holder with stirrer

The STR-773 Water thermostatted cell holder with stirrer is used with a temperature controlled circulating water bath (not supplied) to maintain a constant sample and reference cell temperature.

The STR-773 accepts standard 10 mm path length cells. A magnetic stirrer is included to maintain temperature equilibrium in both cell positions.

Specifications

Cell
10-mm cell
Operating temperature
Room temperature plus 10°C to 90°C
Circulating water inlet tube
Uni-tube (⌽6 x ⌽3 mm)

NCP-706 specifications

Sample movement Linear
Cell type 10-mm rectangular cell, Micro Cell (min. sample width 2 mm)
Thermostatic method Circulation of constant temperature water
Operating temperature Room temperature plus 10° -90°C
Reference cell holder Constant temperature single-cell holder
Dimensions 130 (W) x 255 (D) x 123 (H) mm
Weight 2.5 kg

NCP-705 6-position non-thermostatted automatic cell changer NCP-706 Water thermostatted 6-position automatic cell changer

The NCP-705 and NCP-706 6-position automatic cell changers for up to 6 sample positions. The sample position can be selected using software control and can be used with cell path lengths up to 10 mm.

A single cell is used for the reference position for both the NCP-705/706 automatic six cell changers.

The NCP-706 includes temperature control using a water-thermostatted circulator.

CSP-909 Sample Holder Lid with Syringe Port for V-700 models

The optional CSP-909 lid is used for adding a reagent solution with a syringe without having to open the sample chamber when initiating fast reactions. The syringe port lid can be used together with the STR-773, ETHS-760, ETCS-761 or ETCS-762 cell holders.

Recommended syringe
0507-0220 Micro syringe, 10 µL
0507-0246 Micro syringe, 100 µL

HMC-711 Water Thermostatted Micro Cell Holder (100 µL, 200 µL)

The HMC-711 Water Thermostatted Micro Cell Holder is used with a temperature controlled circulating water bath (not supplied) to maintain a constant sample temperature. The cell holder accepts standard 10 mm path length cells with a minimum path width of 2 or 4 mm.

Specifications

Compatible Cell
5 or 10 mm path length square cell
5 or 10 mm path length square micro cell
Operating temperature
Room temperature plus 10°C to 90°C
Circulating water tube
Uni-tube (⌽6 x ⌽3 mm)

Optional cells

Item P/N Specification
Rectangular micro quartz cell 1103-0025 5 mm path, 2 mm width
Rectangular micro quartz cell 1103-0030 10 mm path, 4 mm width
Rectangular micro quartz cell 1103-0026 10 mm path, 2 mm width
Rectangular micro quartz black cell 1103-0031 10 mm path, 4 mm width
Rectangular micro quartz black cell 1103-0032 10 mm path, 2 mm width

Rectangular micro quartz cell
1103-0026
10 mm path, 2 mm width
100 µL minimum volume

Rectangular micro quartz black cell
1103-0032
10 mm path, 2 mm width
100 µL minimum volume

MHT-745 Manual 4-position water thermostatted turret cell holder

The MHT-745 is used with a temperature controlled circulating water bath. The 4-position turret cell holder can be rotated manually for measurement of up to four samples. Requires four 10×10 mm rectangular cells (not included).

Specifications

Thermostatted cells 10-mm rectangular cell
Sample-side cell holder: 4
Reference side cell holder: 1
Operating temperature Room temperature plus 10 to 90°C
Circulating water inlet Uni-tube (⌽7 x ⌽5 mm)

EHCS-760 Peltier thermostatted single position cell holder (air-cooled)

This Peltier cell holder precisely maintains the specified sample temperature. The cell holder includes a stirrer for uniform cell temperature. The EHC-760 cell holder is an air-cooled accessory for temperature control without using a water circulator.

OPS-512 Optional temperature sensor for EHCS-760

OPS-512 OP sensor for direct temperature measurement in the sample cell for both the ETC-760 or EHC-761.

Specifications

Thermostatted Cell 10-mm rectangular cell (sample side)
Temperature control system Heating/cooling system using Peltier effect
Heat radiating system Air cooled
Stirring system Magnetic stirrer (variable speed)
Temperature setting range 5 to 70°C
Temperature control range 10 to 60°C (for room temperature = 20°C)
Temperature control accuracy ± 0.1°C
Temperature accuracy (Difference between displayed temperature and actual cell temperature)
20°C to 40°C ± 0.5°C
± 1°C when set temperature is outside the above range
Temperature setting difference (Difference between set temperature and displayed temperature)
20°C to 40°C ±0.3°C
± 0.5°C when set temperature is outside the above range

PSC-763 Automatic 6-position Peltier cell changer (air-cooled)

Model PSC-761 multi-cell accessory for simultaneous temperature controlled measurement in up to 6 sample cells. This cell holder uses air-cooled Peltier effect to control sample temperature with high accuracy. Includes a variable speed stirrer for thermal equilibrium for all sample cells.

OPS-513 Optional temperature sensor for PSC-763

OPS-513 OP sensor for direct temperature measurement in the sample cell.

Specifications

Thermostatted Cell Up to six 10 mm rectangular sample cells (ambient reference cell)
Temperature control system Heating/cooling system utilizing Peltier effect
Heat radiating system Forced air cooling
Stirring system Magnetic stirrer (variable speed)
Temperature setting range 10 to 70°C
Temperature control range 15 to 60°C (For room temperature = 20°C)
Temperature control accuracy ±0.1°C
Temperature accuracy (Difference between the displayed temperature and temperature in the cell)
20°C to 40°C ±0.5°C
±1°C when the set temperature is outside the above range
Temperature setting difference (Difference between the set temperature and displayed temperature)
20°C to 40°C ±0.3°C
±0.5°C when the set temperature is outside the above range